Calibration
Calibration Procedure
4-10
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
and press
to confirm the calibration routine has now been
competed.
The Detector Calibration submenu is used to view and set the detector
calibration. The detector calibration screen is visible only when the
instrument is in service mode. For more information on the service
Note
This adjustment should only be performed by an instrument
service technician. Detector Calibration B settings please refer to
Detector Calibration as following description.
s
l
In the Main Menu, choose Service >
Detector Calibration or
Detector Calibration B
.
The Auto Detector Calibration screen allows the user to optimize the
detector performance. This should be done at least once per year and as
often as once per quarter.
l
In the Main Menu, choose Service > Detector Calibration >
Detector Calibration or Detector Calibration B
>
Auto
.
Detector Calibration
Auto Detector Calibration
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......