Servicing
Analog Input Calibration
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
7-21
Use the following procedures to calibrate the analog inputs after
replacing the optional I/O expansion board. These procedures include
selecting analog input channels, calibrating them to zero volts, and then
calibrating them to full-scale using a known voltage source.
Use the following procedure to calibrate the input channels to zero
volts.
1. From the Main Menu, choose Service >
Analog Input Cal
.
The Analog Input Cal screen appears.
Note
If Service Mode is not displayed, refer to “
, then return to the beginning of this step.
s
2. At the Analog Input Cal screen, press
to scroll to a channel,
and press
.
3. With the cursor at Calibrate Zero, press
.
The screen displays the input voltage for the selected channel.
4. Make sure that nothing is connected to the channel input pins and
press
to calibrate the input voltage on the selected channel to
zero volts.
The screen displays 0.00 V as the voltage setting.
5. Press
>
to return to the Analog Input Cal screen and
repeat Steps 2 through 4 to calibrate other input channels to zero as
necessary.
Calibrating the Input Channels to Full-Scale
procedure that follows.
Use the following procedure to calibrate the input channels to full-scale
by applying a known voltage to the channels.
Equipment Required:
DC voltage source (greater than 0 volts and less than 10 volts)
1. Connect the known DC voltage source to the input channel (1-8)
to be calibrated.
shows the analog input pins and
identifies the associated channels.
2. From the Main Menu, choose Service >
Analog Input Cal
.
The Analog Input Cal screen displays input channels 1-8.
3. At the Analog Input Cal screen, press
to scroll to the channel
selected in Step 1, and press
.
Analog Input
Calibration
Calibrating the Input
Channels to Zero Volts
Calibrating the Input
Channels to Full-Scale
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......