Introduction
Principle of Operation
1-4
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
Simultaneous refined mass measurements of sampled particulate on
the filter tape and sample volume measurement provide a continuous
concentration measurement of ambient particulate concentration.
provides a schematic of the components used within the
Model 5028
i
. Within the two measurement chambers the four key
components are 1) the proportional detector, 2) particle collection onto
a glass fiber filter tape, 3) a vertical moving vacuum chamber plate, and
4) the C-14 beta source <100 µCi (<3.7 Mbq). Based on this schematic,
a description of operation is provided.
VACUUM
PUMP
Ta
RHa
Pa
DETECTOR 1
FILTER TAPE 1
TS
VAC
β-SOURCE 1
P
PROPORTIONAL
VALVE 1
TF
HEATER 1
INLET 1
3-WAY V1
1
2
HEATER 2
INLET 2
DETECTOR 2
FILTER TAPE 2
TS
β-SOURCE 2
TF
PROPORTIONAL
VALVE 2
VAC
P
3-WAY V2
1
2
PM2.5 VSCC
Figure 1–1.
Model 5028
i
Flow Schematic
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......