Operation
Service Menu
3-80
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
DETECTOR MANUAL CAL:
>BETA/REF/ALPHA COUNTS
HIGH VOLTAGE
BETA REF THRESHOLD
ALPHA THRESHOLD
RANGE AVG DIAGS ALARM
The Beta/Ref/Alpha Counts screen allows the user to view the real-time
alpha and beta counts.
l
In the Main Menu, choose Service > Detector Calibration >
>
Detector Calibration or Detector Calibration B
> Manual >
Beta/
Ref/Alpha Counts
.
COUNTS:
alpha 0 1/sec
beta 10212 1/sec
beta corr 10145 1/sec
beta ref 5200 1/sec
RANGE AVG DIAGS ALARM
The High Voltage screen allows the user to view and set the high voltage
value.
l
In the Main Menu, choose Service > Detector Calibration >
>
Detector Calibration or Detector Calibration B
> Manual >
High
Voltage
.
HIGH VOLTAGE:
CURRENTLY: 1500 V
SET TO: 1501 V?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Beta Ref Threshold screen allows the user to view and set the
detector operating voltage manually.
l
In the Main Menu, choose Service > Detector Calibration >
>
Detector Calibration or Detector Calibration B
> Manual >
Beta
Ref Threshold
.
REF BETA THRESHOLD:
CURRENTLY: 500
SET TO: 501 ?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Beta/Ref/Alpha Counts
High Voltage
Beta Ref Threshold
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......