Servicing
Flow RTD Replacement
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
7-29
Safety glasses must be worn while replacing the detector assembly.
s
Equipment Damage
Some internal components can be damaged by
small amounts of static electricity. A properly grounded antistatic wrist
strap must be worn while handling any internal component.
s
1. Turn the instrument OFF, unplug the power cord, and remove the
cover.
2. Disconnect the SHV connector from the detector amplifier assembly
and slide throught the wall plate.
3. Loosen both detector mounting screws and slowly lift the detector
assembly.
4. Install the new detector assembly by following the previous steps in
reverse, while taking special care not to puncture the mylar.
5. Calibrate the detector assembly. Refer to the “
procedure on page
Figure 7–16.
Replacing the Detector Assembly
Use the following procedure to replace the flow resistive temperature
diction (RTD) (see
).
Equipment Required:
Sample flow RTD
Adjustable wrench
Equipment Damage
Some internal components can be damaged by
small amounts of static electricity. A properly grounded antistatic wrist
strap must be worn while handling any internal component.
s
1. Turn the instrument OFF, unplug the power cord, and remove the
Flow RTD
Replacement
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......