C-Link Protocol Commands
Alarms
B-16
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
limits in liters per minute/liters per hour. The following example sets the
flow alarm maximum value to 17.34 LPM.
Send:
set alarm flow max a 17.34
Receive:
set alarm flow max a 17.34 ok
Send:
set alarm flow max b 17.34
Receive:
set alarm flow max b 17.34 ok
alarm flow pres min a/b
alarm flow pres max a/b
These commands report the current flow pressure alarm minimum and
maximum value settings. The following example reports that the flow
pressure alarm minimum value is -10.0 mmHg.
Send:
alarm flow pres min a
Receive:
alarm flow pres min a -10.0 mmHg
Send:
alarm flow pres min b
Receive:
alarm flow pres min b -10.0 mmHg
set alarm flow pres min a/b
value
set alarm flow pres max a/b
value
These commands set the flow pressure alarm minimum and maximum
values to
value
, where
value
is a floating-point number representing flow
pressure alarm limits in millimeters of mercury. The following example
sets the flow pressure alarm maximum value to 100.0 mmHg.
Send:
set alarm flow pres max a 100.0
Receive:
set alarm flow pres max a 100.0 ok
Send:
set alarm flow pres max b 100.0
Receive:
set alarm flow pres max b 100.0 ok
alarm flow temp min a/b
alarm flow temp max a/b
This command reports the current flow temperature alarm maximum
value setting. The following example reports that the flow temperature
alarm minimum value is 20.0 °C.
Send:
alarm flow temp min a
Receive:
alarm flow temp min a 20.0 degC
Send:
alarm flow temp min b
Receive:
alarm flow temp min b 20.0 degC
set alarm flow temp min a/b
value
set alarm flow temp max a/b
value
This command sets the flow temperature alarm maximum value
to
value
, where
value
is a floating-point number representing flow
temperature alarm limits in °C. The following example sets the flow
temperature alarm maximum value to 70 °C.
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......