Servicing
Ambient RH/Temp Replacement
7-32
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
4. Verify the sealing o-ring is in position before sliding the new radius
tube assembly into the measurement head.
5. Install the new radius tube assembly by following the previous steps
in reverse.
Figure 7–19.
Replacing the Radius Tube
Use the following procedure to replace the ambient RH/temperature
assembly (
).
Equipment Required:
Ambient RH/temperature assembly
Adjustable wrench
Pliers
Equipment Damage
Some internal components can be damaged by
small amounts of static electricity. A properly grounded antistatic wrist
strap must be worn while handling any internal component.
s
1. Turn the instrument OFF, unplug the power cord, and remove the
cover.
2. Remove the cable from the bottom of the radiation shield assembly
that houses the ambient RH/temperature assembly.
3. Remove the two mounting nuts on the mounting clamp of the
radiation shield assembly. Remove the radiation shield assembly.
4. Turn over the radiation shield assembly and remove the three thumb
screws.
5. Remove the three spacers and mounting bracket.
6. Slide out the ambient RH/temperature assembly.
Ambient RH/Temp
Replacement
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......