Preventive Maintenance
Cleaning the Inlets
5-4
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
Figure 5–2.
European PM
10
Inlet
PM
10
size selective inlet (1 m
3
/h) for 5028
i
:
l
Same construction as the EN reference inlet, but with 1 m³/h
l
With special surface protection using an aluminium EMATAL
surface
l
With special water separator and easy exchangeable impactor baffle
plate
l
Standard tube connection: d = 16 mm for connecting a continuous
monitor
l
Easy converting to PM
2.5
using the PM
2.5
nozzle plate (8 nozzles)
l
Easy converting to PM
1
using the PM
1
nozzle plate (8 nozzles)
If a Sharp-Cut Cyclone (SCC) or a BGI Very Sharp-Cut Cyclone is
being used for PM
2.5
), once every 2-3 weeks the
dust pot should be unscrewed, emptied, and cleaned with a lint-free
cloth. Once every three months, the entire SCC assembly should be
inspected and cleaned, O-rings checked for distortion, cracks, fraying,
or other problems. Replace O-rings as necessary.
Cyclone Maintenance
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......