C-Link Protocol Commands
Hardware Configuration
B-46
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
set temp th a/b
value
This command sets the heater temperature threshold for the sample
to
value
, where
value
is a floating-point number representation of
temperature threshold between 0 to 70 °C. The following example sets
the temperature threshold to 35 °C.
Send:
set temp th a 35
Receive:
set temp th a 35 ok
Send:
set temp th b 35
Receive:
set temp th b 35 ok
contrast
This command reports the screen’s level of contrast. The following
example shows the screen contrast is 55%, according to
Send: contrast
Receive:
contrast 11: 55
%
set contrast
level
This command sets the screen’s level of contrast, according to
.
The following example sets the contrast level to 50%.
Send:
set contrast 10
Receive:
set contrast 10 ok
Table B–6.
Contrast Levels
Level
Contrast Level
0
0%
1
5%
2
10%
3
15%
4
20%
5
25%
6
30%
7
35%
8
40%
9
45%
10
50%
11
55%
12
60%
13
65%
14
70%
15
75%
Hardware
Configuration
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......