C-Link Protocol Commands
Alarms
B-12
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
Send:
alarm alpha count min a
Receive:
alarm alpha count min a 200
Send:
alarm alpha count min b
Receive:
alarm alpha count min b 200
set alarm alpha count min a/b
value
set alarm alpha count max a/b
value
These commands set the alpha count alarm minimum and maximum
values to
x
, where
x
is an integer representing alpha count limits. The
following example sets the alpha count alarm maximum value to 500.
Send:
set alarm alpha count max a 500
Receive:
set alarm alpha count max a 500 ok
Send:
set alarm alpha count max b 500
Receive:
set alarm alpha count max b 500 ok
alarm amb rh min
alarm amb rh max
These commands report the current ambient relative humidity alarm
minimum and maximum value settings. The following example reports
the ambient relative humidity alarm minimum value is 35.0%.
Send:
alarm amb rh min
Receive:
alarm amb rh min 35.0
%
set alarm amb rh min
value
set alarm amb rh max
value
These commands set the ambient relative humidity alarm minimum
and maximum values to
value
, where
value
is a floating-point number
representing ambient relative humidity alarm limits in percent. The
following example sets the ambient relative humidity alarm maximum
value to 55%.
Send:
set alarm amb rh max 55
%
Receive:
set alarm amb rh max 55
%
ok
alarm amb temp min
alarm amb temp max
These commands report the current ambient temperature alarm
minimum and maximum value settings. The following example reports
that the ambient temperature alarm minimum value is 4.0 °C.
Send:
alarm amb temp min
Receive:
alarm amb temp min 4.0 degC
set alarm amb temp min
value
set alarm amb temp max
value
These commands set the ambient temperature alarm minimum and
maximum values to
value
, where
value
is a floating-point number
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......