Operation
Service Menu
3-78
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
The Auto Mass Coefficient screen allows the user to conduct a mass foil
calibration.
Note
After pressing
a series of screens will walk you through the
mass foil calibration procedure. For details, see “
Bookmark not defined.
s
l
In the Main Menu, choose Service > Mass Calibration >
Mass
Calibration or Mass Calibration B>
Mass >
Auto
.
MASS COEF: 7000.0
FOIL VALUE: 144
2
3
µ
g
TO SAVE FOIL VALUE
AND OPEN BENCH
RANGE AVG DIAGS ALARM
The Manual Mass Coefficient screen allows the user to adjust the mass
coefficient used in beta attenuation.
l
In the Main Menu, choose Service > Mass Calibration >
Mass
Calibration or Mass Calibration B>
Mass >
Manual
.
MASS COEFFICIENT:
CURRENTLY: 7000
SET TO: 699
2
?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Alpha Efficiency Coefficient screen allows the user to view and set
the alpha efficiency value.
Note
This value is set in the factory and should not be adjusted.
s
l
In the Main Menu, choose Service > Mass Calibration >
Mass
Calibration or Mass Calibration B>
Alpha Eff Coefficient
.
ALPHA EFFICIENCY COEF:
CURRENTLY: 0.120000
SET TO: 0.12000
1
?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
Auto Mass Coefficient
Manual Mass Coefficient
Alpha Efficiency Coefficient
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......