Servicing
Themistor Replacement
7-22
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
4. Press
to scroll to Calibrate Full-scale, and press
.
The screen displays the input voltage for the selected channel.
5. Use
and
to enter the source voltage, and
press
to calibrate the input voltage for the selected channel
connected to the source voltage.
6. Press
>
to return to the input channels display and
repeat Steps 3-5 to calibrate other input channels to the source
voltage as necessary.
Use the following procedure to replace the optional ambient
temperature thermistor (
Equipment Required:
Thermistor assembly
Equipment Damage
Some internal components can be damaged by
small amounts of static electricity. A properly grounded antistatic wrist
strap must be worn while handling any internal component.
s
1. Turn the instrument OFF, unplug the power cord, and remove the
cover.
2. Squeeze the thermistor latch and pull the thermistor assembly from
the AMB TEMP connector on the measurement interface board.
3. Snap the new thermistor into the AMB TEMP connector.
Figure 7–10.
Replacing the Thermistor
Themistor
Replacement
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......