Operation
Service Menu
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
3-81
The Alpha Threshold screen allows the user to view and set the alpha
threshold manually.
l
In the Main Menu, choose Service > Detector Calibration >
>
Detector Calibration or Detector Calibration B
> Manual >
Alpha
Threshold
.
ALPHA THRESHOLD:
CURRENTLY: 1500
SET TO: 1501 ?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Analog Output Calibration menu provides access to the 6 voltage
channels and 6 current channels for calibration. Current channels are
visible only if the I/O expansion board is installed. The analog output
calibration menu is visible only when the instrument is in service mode.
For more information on the service mode, see “
in this chapter.
Note
This adjustment should only be performed by an instrument
service technician.
s
l
In the Main Menu, choose Service >
Analog Out Calibration
.
ANALOG OUTPUT CAL:
>VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
RANGE AVG DIAGS ALARM
ANALOG OUTPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM
VOLTAGE CHANNEL 6
CURRENT CHANNEL 1
Alpha Threshold
Analog Output Calibration
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......