Servicing
Front Panel Board Replacement
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
7-33
7. Slide in the new ambient RH/temperature assembly by following
the previous steps in reverse.
8. Calibrate the instrument. Refer to the “
manual.
Unfasten Thumb Screws (3)
Remove Cable
Remove RH/Temperature
Assembly
Figure 7–20.
Replacing the Ambient RH/Temperature Assembly
Use the following procedure to replace the front panel board (
).
Equipment Required:
Front panel board
Equipment Damage
Some internal components can be damaged by
small amounts of static electricity. A properly grounded antistatic wrist
strap must be worn while handling any internal component.
s
1. Turn the instrument OFF, unplug the power cord, and remove the
cover.
2. Remove the three ribbon cables and the two-wire connector from
the front panel board.
3. Pop the board off of the two top mounting studs and remove the
board by lifting it up and off the slotted bottom support.
4. Replace the front panel board by following the previous steps in
reverse.
Front Panel Board
Replacement
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......