Operation
Instrument Controls Menu
3-22
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
The Standard Pressure screen is used to set the standard pressure.
Note
Pressure B settings please refer to Pressure as following
description.
s
l
In the Main Menu, choose Instrument Controls > Volumetric
Conditions > Pressure or Pressure B >
Standard
.
STANDARD PRESSURE:
CURRENTLY: 760 mm
Hg
SET TO: 760 mm
Hg ?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The iSeries instruments include a built-in datalogging capability as a
standard feature. The operator is allowed to create two different types
of records, which for historical reasons, are named lrecs and srecs. Each
record can contain up to 32 different fields, or data items, and records
can be created at user-defined intervals ranging from 1 to 60 minutes.
Record generation is tied to the instrument’s real-time clock. For
example, if the logging period for srecs is set to 30 minutes, a new srec
will be generated on every hour and every half hour (10:00, 10:30,
11:00 …). Lrecs and srecs function independent of each other. So, for
example, an srec containing just the current PM level could be generated
every 5 minutes while an lrec containing a full set of diagnostic data
could be generated once every hour.
The analyzer’s computer system includes three megabytes of flash
memory which is enough to store a full lrec containing 32 data items
and a full srec containing 32 items once each minute for a week (>20,000
total records). If logging is limited to the minimum content of date,
time, PM concentration and error flags, the analyzer can store data once
each minute for 4 months (>190,000 total records).
The Datalogging Settings menu allows the user flexibility in how data is
stored and recorded.
l
In the Main Menu, choose Instrument Controls >
Datalogging
Settings
.
DATALOGGING:
>SELECT SREC/LREC SREC
VIEW LOGGED DATA
ERASE LOG
SELECT CONTENT
COMMIT CONTENT
RANGE AVG DIAGS ALARM
RESET TO DEFAULT CONTENT
CONFIGURE DATALOGGING
Standard
Datalogging Settings
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......