C-Link Protocol Commands
Measurements
B-10
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
avg24 time a/b
This command reports the 24 hour averaging time in hours and
minutes. The following example shows that the averaging time is set to
18:30.
Send:
avg24 time a
Receive:
avg24 time a 18:30
Send:
avg24 time b
Receive:
avg24 time b 18:30
set avg 24 time a/b
hh:mm
hh
= hours (01 to 23)
mm
= minutes (01 to 59)
This command sets the 24 hour averaging time to start averaging the
concentration from hh:mm until next 24 hours. It is then set to next 24
hours automatically. The following example sets the 24 hour averaging
time to 17:50.
Send:
set avg24 time a 17:50
Receive:
set avg24 time a 17:50 ok
Send:
set avg24 time b 17:50
Receive:
set avg24 time b 17:50 ok
flags
This reports 8 hexadecimal digits (or flags) that represent the status of
the mode, pressure and temperature compensation status, concentration
units, concentration modes and alarms. To decode the flags, each
hexadecimal digit is converted to binary as shown in the
. It
is the binary digits that define the status of each parameter.
Send:
flags
Receive:
flags 00000000
Measurements
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......