AFA4000/2
Airflow Monitor / p.131
Menu item
Option
Instructions
Sash 1 settings
(cont.)
Auto Lock
Enables / disables the Auto-Lock
feature. See section 7.4.2.
Lock Alarm
Sets the alarm delay time. This is the
time before alarm activates if sash
remains locked.
(1 to 30 minutes)
Lock Pos Calibration
Calibrates the Lock position. See
section 7.4.3.
Sensitivity
Set the sensitivity from 0
–
6, where 6
is the least sensitive. This allows the
sash travel to stutter for 1 second
before the Sash Fault alarm is
triggered.
8.3.2.3
BMS input settings
Menu item
Option
Instructions
BMS Input Settings Fire Alarm
Set the Fire Alarm BMS input:
•
Disabled
•
Input 1
–
4 open
•
Input 1
–
4 closed
Open Sash
Set the Open Sash BMS input:
•
Disabled
•
Input 1
–
4 open
•
Input 1
–
4 closed
Close Sash
Set the Close Sash BMS input:
•
Disabled
•
Input 1
–
4 open
•
Input 1
–
4 closed
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...