AFA4000/2
Airflow Monitor / p.100
This instructs the controller which direction the motor needs to turn to close the sash.
12.
Move the sash to the lowest position required, press
Enter
.
13.
Either:
•
Press
Mute,
if this is the only pre-set position required, or
•
Set the sash to the normal working height, for example 500 mm.
14.
Press
Enter
, for two position operation.
15.
Either:
•
Press
Mute
, if only two pre-
set positions (‘lowest’ and ‘normal’) are required.
•
Set the sash to the full opening height.
16.
Press
Enter
, for three position operation.
The lowest, normal and fully open positions are now calibrated.
17.
Press
Enter
, to exit Calibration mode.
18.
Press
↓
button, select
Back to Set up
menu.
19.
Press
Enter
.
20.
Press
↓
button, select
Exit and Save Changes.
21.
Press
Enter
.
The Controller will now enter
Run
mode.
Note
:
When the Auto Sash motor drive is set to a high speed, allow a margin for the sash
to over-run when calibrating the sash position. See the instructions above. Allow a
5 mm (0.197") margin at the highest and lowest point.
7.4.3
Calibrating the sash lock function
To define the required Sash Lock position:
1.
Press and
hold the
Enter
button for 5 seconds until the
Main Menu
is displayed.
2.
The current Hardware and Software versions are displayed, for example
1. HW2.0
FW1.1
3.
Press
↓
button, select
2. Setup
and press
Enter
.
4.
Enter the Password using the
↑
arrow and
Enter
button, the factory default is 1 0 0 1.
In the Set-Up menu, the controller will display
Keypad Tones
.
5.
Press
↓
button, select
Sash 1 Settings
and press
Enter
. The controller will display
1. S1 Enabled.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...