AFA4000/2
Airflow Monitor / p.63
Figure 33: Front view of rack and pinion sash drive
6.
When fitting a sash low switch instead of a sash position sensor, the rack will need to
be cut to the correct length to enable the switch to operate.
Close the sash fully and mark the rack just below the position of the proximity low
switch at the top of the motor assembly to ensure that the rack is clear of the switch
when the sash is closed.
7.
Connect the Sash Drive to the Auto Sash Control Unit using the cables provided (with
red and black connectors). See the wiring schematic in section 6.10 and Figure 31.
Note: It is important to observe the correct polarity when making electrical
connections.
8.
Connect the clutch assembly to the Auto Sash Control Unit using the cables provided
(with blue and yellow connectors).
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...