AFA4000/2
Airflow Monitor / p.82
Figure 55: Sash position sensor draw wire attached by cable tie
When the sensor is fitted, manually open / close the sash to ensure that the sensor is
installed correctly.
6.9
Personnel sensor
The Personnel Sensor is supplied as a single transmitter / receiver unit.
6.9.1
Installing the personnel sensor
1.
Use the mounting template illustrated in Figure 56 and the fixing screws provided to
mount the Personnel Sensor. Mount the sensor centrally to the front of the fume
hood and above the fume hood sash, use the fixing holes in the back of the sensor.
See the table in Figure 56 for the correct mounting hole dimensions and locations.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...