Installing and aligning the Under Sash sensor
The Under Sash Sensor consists of two components:
•
Sash Sensor
•
Retro-reflective Tape Strip
1.
Fit the sensor so that its lens is clear of obstructions and below the bottom of the
sash handle.
Figure 46: Under sash sensor mounting
Figure 47: Under sash sensor dimensions
Use a suitable bracket so that the sensor position can be adjusted to the required
position.
2.
Fix the 5 m cable, which is attached to the sensor, to the sash wire using cable ties.
Attach the 5 m cable to the sash counterweight to keep the cable taught as the sash
opens or closes.
Encase the cable from the sash counterweight to the sash controller box can in plastic
sheathing to guide the cable slack and stop it from snagging.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...