AFA4000/2
Airflow Monitor / p.91
6.11
Start up
Make sure the Auto Sash Controller is installed before attempting to calibrate the
controller.
When the unit is powered up, the following events occur:
The alarm performs a self-test of its functions, LEDs and audible alarm, which takes
approximately 3 seconds. All inputs and outputs are deactivated during the self-test.
At the end of the delay, the unit will do one of two things:
•
When the controller has been calibrated, it enters normal operating mode and
displays the current status or alarm condition.
•
When the controller has not been calibrated, the monitor displays the messages:
•
Sash drive inhibited
, if the tilt switch input is open
•
Auto Sash Not Configured
, if the tilt switch is closed
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...