AFA4000/2
Airflow Monitor / p.135
Setting up the temperature input:
1.
From the
Requires set up
screen, press
Mute
, or from the
Run Screen
press and hold
the
Mute
button for 5 seconds until the
Main Menu
is displayed.
2.
Use the
↑
/
↓
buttons to select
Set Up Monitor > Configure
.
3.
Enter the password (the factory default password is 0-0-0-0) and press
Mute
to
continue.
4.
Use the
↑
/
↓
buttons to select
Input 3
>
Analog
> Temperature
.
5.
Use the
↑
/
↓
buttons to select the
Temperature
menu.
Parameter
Description
Range
A Temperature Units
Display units
°C or °F
B Low Temperature Alarm Sets the Low Temperature Alarm
point
0.0 to High Alarm
value
C High Temperature
Alarm
Sets the High Temperature Alarm
point
100.0 to Low Alarm
value
D Show Temperature
Turns the temperature display
On / Off
Yes / No
E Offset
Allows the measured value to be
offset
-5.0 °C to +5.0 °C
Note: The temperature sensor is accurate to 0.5 °C. The offset parameter can be used,
if the displayed value needs to be changed to match a 3rd party instrument or
other equipment.
6.
Use the
↑
/
↓
buttons to select
Temperature Units
, use the
↑
/
↓
buttons to select
Centigrade / Fahrenheit
.
7.
Use the
↑
/
↓
buttons to select
Low Temperature Alarm
, use the
↑
/
↓
buttons to set
the alarm point.
8.
Use the
↑
/
↓
buttons to select
High Temperature Alarm
, use the
↑
/
↓
buttons to set
alarm point.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...