Airflow Monitor / p.83
A
Cable Entry
10 x 20 mm (5 mm diameter cable provided)
B
Fixing Hole
5 mm diameter maximum
C
Spare Hole
10 x 20 mm
D
Spare Hole
10 x 20 mm
E
Fixing Hole
5 mm diameter maximum
F
Spare Fixing Hole
5 mm diameter maximum
Figure 56: Personnel sensor mounting details
The Personnel Sensor should be positioned so that it points straight down, with the
horizontal axis of the sensor parallel to the horizontal axis of the fume hood (Figure 57).
Note: Do not mount the sensor higher than 250
0mm (98”) fro
m the laboratory floor.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...