Figure 49: Maximum under sash sensor offset
For example, on a 1000 mm wide Fume Hood the sensor can detect the retro-
reflective strip up to ± 20 mm either side of the Under Sash Sensor's centreline.
4.
Connect to the Auto Sash Control Unit. See the wiring schematic in section 6.10 and
Figure 31.
6.6
Tilt switch
Tilt Switches are provided in a variety of forms, the installation instructions that follow
are therefore general.
•
Ensure that the Tilt Switch is installed in the correct orientation
–
‘open’ when the
front panel is closed and ‘closed’ when the front panel
is open.
•
Connect the Tilt Switch to the Auto Sash Control Unit. See the wiring schematic in
section 6.10 and Figure 31.
•
Link out / connect the relevant terminals on the Auto Sash Control Unit when not
using a Tilt Switch. See the wiring schematic in section 6.10 and Figure 31.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...