AFA4000/2
Airflow Monitor / p.79
Figure 51: Sash low micro-switch wiring schematic
6.7.2
Installing the Sash Low proximity switch
The Sash Low Proximity Switch is an inductive switch that operates when it comes within
range of a metal plate attached to the sash, i.e. when the sash is closed.
The type of switch selected makes or breaks the circuit when the sash is closed. Typically,
the switch would be fitted so that the switch is closed when the sash is open.
Connect the Sash Low Proximity Switch to the Auto Sash Control Unit (Figure 52). See
the wiring schematic in section 6.10 and Figure 31.
Figure 52: Sash low proximity switch wiring schedule
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...