Figure 53: Sash Low Proximity Switch Dimensions
Note: When the Auto Sash control is being retro-fitted, fix the bracket to allow for over-
travel of the sash.
6.8
Sash position sensor
The sash position sensor is fitted with a mounting bracket which can be fitted in four
different orientations to enable the sensor draw wire to be correctly aligned.
1.
Offer up the sensor to determine the best fixing position on the fume hood
framework.
Where possible, ensure the sensor draw wire runs straight out of the sensor housing,
in line with the cable housing extension. Where the wire runs at an angle to the
sensor, ensure the angle is as small as possible to prevent the wire from rubbing on
any part of the sensor body, including the cable housing extension.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...