Overview: Auto Sash Controller
The Auto Sash Controller automates the opening and closing of the fume hood sash.
Automation of the sash movement provides an increased level of safety to the controlled
environment. A sash cannot be accidentally left open and can respond to safety critical
inputs from external sources, such as fire alarms and ventilation system failures.
The Auto Sash Controller facilitates cost savings. When linked to ventilation controls, the
automatic closing of an unattended fume hood allows extractor fan speeds and
electricity consumption to be reduced.
4.1
Features
The Auto Sash Controller has the following features:
•
Auto-Close
The sash automatically closes when the fume hood is
unattended.
•
Auto-Open
The sash automatically opens when the user returns.
•
Sash Lock
The sash is programmed to lock at a designated
position.
•
Manual Sash
Open/Close
To open/close the sash manually.
•
Keypad Pushbutton
Open/Close
To open/close the sash using the Keypad Pushbuttons.
•
Footswitch
To open / close the sash using an optional, third party
supplied, footswitch.
•
Tilt Switch
To disable the Auto Sash Control Unit using a, third
party supplied, tilt switch.
•
Building Management
System (BMS) Inputs
The sash automatically closes when a signal is received
from the BMS.
•
Building Management
System (BMS) Outputs
The Auto Sash Controller sends signals to the BMS
regarding the current status of the fume hood.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...