AFA4000/2
Airflow Monitor / p.122
Menu item
Option
Instructions
Input / Output
Diagnostics
> TLP POS F R
S1 X-X 0.0 - -
Sash 1 position.
An active output is shown with an X
beneath it. In the example shown here,
the tilt switch (T) and personnel sensor
(P) are active and the light curtain (L) is
inactive as shown by the X-X on the
second row.
Parameters are:
T
Tilt switch
L
Light curtain
P
Personnel sensor
POS Sash position. When in use,
the value goes from 0 (sash
closed) to 100% (sash fully
open) to show the position of
the sash.
F
When
Mute
is pressed, an X
appears in the bottom row to
indicate that the motor is
driving forward.
R
When
Mute
is pressed, an X
appears in the bottom row to
indicate that the motor is
driving in reverse.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...