AFA4000/2
Airflow Monitor / p.68
When the motor drive has been installed, open / close the sash to ensure that the sash
can travel freely.
6.4
Sash light curtain
The Sash Light Curtain consists of three main components and associated wiring:
•
Sash Light Curtain Control Box
•
LED Transmitter
•
LED Receiver
The Sash Light Curtain Control Box is connected to the main Auto Sash Control Unit.
6.4.1
Installing the sash light curtain control box
1.
Fix the Sash Light Curtain Control Box onto the top of the fume hood, using the tabs
attached to the control box and the screws provided. Ensure that there is access to
the terminals to connect the wiring.
Figure 38: Sash Light Curtain Control Box
2.
Use the pre-assembled cables and connectors supplied to connect the Light Curtain
Control Box to the Auto Sash Control Unit. See the wiring schematic in section 6.10
and Figure 31.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...