AFA4000/2
Airflow Monitor / p.99
2.
At the end of the self-test, the Auto Sash Control Unit will do one of two things:
If the Auto Sash Control Unit has previously been calibrated:
•
The Auto Sash Controller enters the Normal Operating Mode and displays its
current status or alarm condition.
If the Auto Sash Control Unit has not been calibrated:
•
When the Tilt Switch input is open, the display shows
'Sash Drive Inhibited
'.
•
When the Tilt Switch input is closed, the display shows
‘
Auto Sash Not Configured
'.
7.4.2
Calibrating the Auto Sash Control Unit for Tiptronic operation using the
Auto Sash keypad
The Auto Sash Control Unit must have each of its required stationary positions defined
for Tiptronic functionality:
1.
Press and hold the
Enter
button for approximately 5 seconds or until the main menu
appears.
The LCD display will show the current hardware and software versions. For example,
1.
HW2.0 FW1.1.
2.
Press
↓
and select
2. Setup.
3.
Press
Enter
.
4.
Enter the Password using the
↑
arrow and
Enter
button, the factory default is 1 0 0 1.
In the Set-Up menu, the controller will display
Keypad Tones
.
5.
Press
↓ and
select
Sash 1 Settings.
6.
Press
Enter
.
In the S1 (Sash 1) Settings menu, the display will show
1. Enabled.
7.
Press
↓
and select
8. Calibration.
8.
Press
Enter
.
The display will show on screen instructions. For example:
"Manually move sash to centre and then press the Up/Down keys to find sash down
direction. Press Cancel to abort Calibration"
9.
Move the sash to the middle of the sash opening.
10.
Press
↑/↓
buttons to determine which arrow activates sash closing.
11.
Press the
↑/↓
button which closes the sash, press
Enter
.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...