If the glass object does not break the beam, adjust the gain potentiometer counter-
clockwise until inserting the glass object breaks the beam.
15.
If required, set
sw4
to
Off
to disable the buzzer.
If the sash light beams intermittently sense the glass object, ensure
sw1
is set to Off
(short range) and repeat steps 5 to 14 above.
If the beam is broken when the path is clear and adjusting the Gain Potentiometer
does not resolve this situation, either:
Switch
sw2
On
(Diagnostic mode). After approximately 15 seconds the
RX ERR
and
TX
ERR
LEDs (Figure 65) will indicate if there is a faulty beam:
•
RXERR Red LED
On
–
Faulty receiver
•
TXERR Red LED
On
–
Faulty transmitter
Or:
Clean or replace the faulty component and set
sw2
to
Off
(Operation mode).
7.4.5
Tilt switch
No user calibration of the Tilt Switch is possible. When using a Tilt Switch, ensure that
the top panel of the fume hood is closed before switching on the Auto Sash Control
Unit.
7.4.6
Sash position sensor
No user calibration of the Sash Position Sensor is possible. The sensor provides sash
position data to the Auto Sash Controller, relative to its starting position.
7.4.7
Personnel sensor
The sensor is self-calibrating and does not need adjusting.
For optimum performance, ensure the detection field in front of the fume hood is clear
from objects and personnel when powering up the Auto Sash Control Unit.
The sensor will be fully operational approximately 10 seconds after the sensor is
powered up.
The LED on the sensor will be green if the detection area is clear and red if it is not clear.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...