AFA4000/2
Airflow Monitor / p.61
•
The installation is part of a new build or retro-fit.
•
The size and configuration of the fume hood.
•
The fume hoods location.
None of these options preclude manual operation of the sash.
6.3.1
Rack and pinion motor drive
The Rack and Pinion Motor Drive uses a rack fitted directly onto the sash frame and a
motor drive assembly fitted to a framework above the fume hood.
This drive system is ideal for retro-fit installations where there is no access to the sash
wires or counterweights and / or where the sash is fitted to a metal framework, for
example combination type sash fume hoods.
6.3.1.1
Installing the rack and pinion drive
1.
Raise the sash to the fully open position.
2.
Offer up the motor assembly and rack to ensure that there is a clearance of at least
1.2 m (47.24") above the top of the sash.
Note: The rack is supplied in 1m (39.37") lengths
3.
The motor drive assembly should be fitted to a rigid framework and positioned
centrally above the sash. To ensure that the T-bar fixing clears the bottom of the
motor assembly, there should be at least 50 mm between the top of the sash and the
bottom of the motor drive assembly, when the sash is fully opened.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...