AFA4000/2
Airflow Monitor / p.88
6.9.4
Personnel sensor dip switch settings
The following Personnel Sensor functions and parameters are controlled by Dip switches,
located within the Sash Light Curtain Control Box (Figure 62) (See section 7.4.7).
Figure 62: Personnel Sensor Dip Switch Settings
Switches
Setting
Description
1 & 2
Presence timer
Select the time delay before the Personnel Sensor
‘re
-
learns’ the background whe
n a stationary
object is detected.
2, 15, 60 or 180 second delays are selected by
setting associated switches 1 and 2 to
combinations of ON/OFF (Figure 62).
Note: The timer will reset if any movement is
detected during the time period.
3 & 4
Pattern depth
Set the distance of the detection field away from
the sensor.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...