AFA4000/2
Airflow Monitor / p.117
Menu item
Option
Instructions
Sash 1 settings
(cont.)
Auto Clear Obst Alarm
Select whether the system auto resets
the Obstruction Alarm, if Obstruction is
removed.
Personnel
Set the input activation for the
personnel sensor:
•
Normally Open (N/O) Contact
•
Normally Closed (N/C) Contact
Disabled
Personnel Power Cycle
Follow the on-screen instructions to
manually calibrate the personnel
sensor, using the menu.
Calibration
Follow the on-screen instructions to
manually calibrate the sash position
sensor and motor direction.
Close Threshold
Set the height that the Auto Sash will
accept as closed if the sash is stopped
during travel.
(0
–
100mm)
Tiptronic
Select
Enable
to open or
Disable
to
close the Tiptronic touch sash.
Min TipTime
Set the tiptronic touch sensitivity
minimum time.
(100
–
300 ms)
Max TipTime
Set the Tiptronic touch sensitivity
maximum time.
(500
–
1500 ms)
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...