AFA4000/2
Airflow Monitor / p.120
Menu item
Option
Instructions
BMS Output
Settings
Alarm Sounder
Enable / disable the alarm sounder.
Analog Volts
Select the required analogue voltage
output:
•
0
–
10V
•
1
–
10V
•
0
–
5V
•
1
–
5V
Analog Current
Select the required analogue current
output:
•
0
–
20mA
•
4
–
20mA
Relays 1 to 4:
RLY1, RLY2, RLY3, RLY4
Select the required settings for the
relay:
Note: Each of the four relays have the
same setting options.
•
Always Open
•
Always Closed
•
Open on Sash Alarm
•
Close on Sash Alarm
•
Open on Obstruction
•
Close on Obstruction
•
Open on Personnel
•
Close on Personnel
•
Open on Sash Open
•
Close on Sash Open
•
Open on Sash Locked
•
Close on Sash Locked
Exit and Abandon
Changes
Return to Main
without saving
any changes.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...