How to Set Up Your GX Series Microlog
Global Measurement Settings
measurement following a non-ICP measurement). Thereafter, data acquisition times
are unaffected. If ski slope problems occur, try increasing the settling delay by one or
two seconds.
Route ICP
–
Never
,
As Database
, or
Auto
matic. Most sensors contain an internal
amplifier that requires ICP, whereas some sensors do not use ICP. Determine your
ROUTE sensor’s needs and disable or enable ICP by selecting
Never
,
As Database
, or
Auto
matic. If the ROUTE ICP is set to
As Database
, the
Enable ICP
setting in @ptitude
Analyst POINT Properties is used to determine whether ICP is used for each POINT; if
set to
Never
, the ICP will never be used; if set to
Automatic
and the
Enable ICP
setting
in @ptitude Analyst POINT Properties is enabled, the ICP is ON if the transducer type is
Accelerometer, and is OFF for all other transducer types (even if
Enable ICP
is enabled).
The ICP indicator displays in the title bar if the se
lected POINT is to be taken with ICP
power ON.
Figure 2 - 3.
A Data Collection
Screen with ICP Indicator On.
View MPA Spectrum
– Select whether to display the MPA spectrum for
First
,
Last
,
All
,
or
None
. The GX displays the corresponding spectrum (including any FAM or spectral
band information) allowing the operator to inspect and verify that the data is correct.
First
–
Displays the first MPA POINT
in a spectrum.
Last
– Displa
ys the last MPA POINT in a spectrum.
All
–
Displays all MPA POINTs in a spectrum.
None
– Does not display a spectrum.
Trigger – Manual
or
Visual
. (Available only if the
Setup
screen’s
Trigger
field is set to
User
.) Determines how the trigger level is entered when setting up measurements that
require a trigger, either manually as numeric input in the subsequent
Trigger Level
field, or visually using a graphic representation of the trigger signal and trigger level.
Manual
–
When s
et to
Manual
, you will enter the trigger level in volts as
numeric input in the
Trigger Level
setup field.
Visual
– Displays the
Set Trigger
screen that shows a graphical representation
of the tacho signal and trigger level (horizontal cursor). Use the up and down
arrow keys to move the trigger cursor to the desired trigger level (typically
between ½ and ¾ of the full trigger signal), and press
OK
. The
Set Trigger
screen closes, the
Trigger
setting changes to
Manual
and the visually specified
trigger level numerically displays in the
Trigger Level
field.
Trigger Level –
Data entry field for manual trigger level input, or displays resulting
trigger level when the trigger level is entered visually (see above
Trigger
explanation).
SKF Microlog - GX Series
2 - 9
User Manual