How to Set Up Your GX Series Microlog
System Settings
Audio Output Status – Normal Display
,
Auto Hide Display
, or
Control Off
.
Normal Display
– The audio output volume always appears in the title bar of
applicable data collection screens
Auto Hide Display
– The audio output volume appears in the title bar of
applicable data collection screens only while one is adjusting the volume
Control Off
– Volume is not adjustable
Transducer Change – Do not warn
or
Warn
. Determines whether a warning message
appears on the Microlog’s display when the next measurement uses a different ICP
setting than the previous measurement. This warning helps make the data collection
person aware that the upcoming measurement requires a sensor change (e.g.,
temperature sensor to vibration sensor, or vice-versa).
Bias Check – Manual
or
Automatic
Manual
-
Within the ROUTE hierarchy, press Shift+B to check the bias voltage
for ICP measurements. This turns on the ICP supply (if it wasn't already) and
measures the DC bias voltage for each channel used by this measurement
(CH1 for single channel POINTs, typically CH1 and CH2 for two
-channel
POINTs, and CH1, CH2, and CH3 for Triax POINTs). A Bias check results
screen displays.
This screen shows the voltage(s) for each channel, along with a red / green
indication: green if the bias voltage is between 2 V and 18 V, which is regarded
as normal, and red if it is outside this range.
This can be useful to troubleshoot a potential cable or accelerometer problem.
Automatic
– Automatically checks sensor bias at the start of each
measurement.
Nonroute
–
Use Analyzer Module
or
Use Legacy Nonroute
. Determines which
nonroute mode is available fr
om within a ROUTE. From either the Analyzer module or
the legacy Nonroute mode, you can collect measurement data for POINTs not
previously downloaded to the Microlog from your @ptitude Analyst host software. Data
collected in either mode can be subsequently uploaded to @ptitude Analyst as nonroute
data.
Language
– Use the drop down list to select the language for GX Series Microlog
displays. After entering your selection, press
Apply
to return to the main screen, all
display text appears in the specified language.
•
After specifying changes to
Setup
screen settings, press the
Apply
function button
to store your new settings in the GX Series Microlog. The main screen displays.
2 - 6
SKF Microlog - GX Series
User Manual