Appendix D
Safety Instructions
Instrument Installation Drawing (090-22600-1 Rev B)
Hazardous (Classified) Location Class I, Division 2, Groups A, B, C,
and D
1.
Suitable for use in Class I, Division 2, Groups A, B, C, D or non-
hazardous locations only
2.
WARNING
-Explosion hazard-batteries must be changed in an area
known to be non-hazardous
3.
WARNING
-Explosion hazard-do not connect/disconnect while circuit
is live unless area is know to be non-hazardous
4.
WARNING
-Explosion hazard-substitution of components may
impair suitability for Class I, Division 2
5. Unused connectors on the instrument must be covered by sealing
caps, p/n: 012-10372 or 012-10776
6. The SD card shall not be disconnected in hazardous area. It must be
installed in safe area only
7.
Only acceler
ometers listed in table 1 may be used with the
instrument in a hazardous area
Accessories
Only the following accessories may be used with the instrument in the hazardous area.
Accelerometers
Manufacturer
Model
SKF Badged
Wilcox
on
762-33
CMSS762-CA
734-33
766-33
CMSS 766-CA
793-
33, 793L
-33
CMSS 793-CA, CMSS
793L
-CA
793V-33, 793V-5-33
CMSS 793V-CA, CMSS 793V-5-CA
797-
33, 797L
-33
CMSS 797-CA, CMSS
797L
-CA
993B-x-35
IMI
CS622A01
CS625B01
Batteries
Use BP-12 only.
The BP-12 battery can be charged out with the hazardous location using a power
supply supplied by the manufacturer and CA-38 connected to the instrument.
SKF Microlog - GX Series
D - 1
User Manual