Non ROUTE
How to Collec
t Pre-
set Non ROUTE Measurements
OR
•
Press the
Type In
button to manually enter the running speed using the numeric
keypad.
•
Press the
OK
function button to save the specified RPM value and begin data
collection, or press the
Cancel
button to discard changes.
OR
•
Press the
Spec
button to manually enter the RPM value from the spectral display.
After autoranging and data acquisition is complete, the
Enter RPM
display appears.
Figure A - 4.
The
Enter RPM
Plot Display Screen.
cursor value
•
Use the arrow buttons to navigate the cursor over the spectrum to indicate the 1X
RPM (1X running speed).
Use the (
P
) key to quickly jump from peak to peak.
•
Press the
OK
function button when the cursor is located on the spectrum’s 1X
frequency. Data collection begins.
Collecting and Displaying Orbit Data
After pressing an Enter button from the Orbit Setup screen, or after specifying the
measurement’s RPM value, data collection takes place and the orbit display appears.
A - 8
SKF Microlog - GX Series
User Manual