3152A User Manual
Using The Instrument 3-8
Racal Instruments
©
2001
/* Change the output waveform to square, frequency to 2 MHz,
amplitude to 5 V, offset to 2.5 V and duty cycle to 25%. Note that
there is no need to use the FUNC:MODE command because the
default value after RESet is FIXed.*/
:APPLy:SQUare
2e6,5,2.5,25
/* Turn the main output on.*/
:OUTPut
ON
/*Turn the SYNC output on, if required. */
:OUTPut:SYNC
ON
If the above commands are executed correctly, a square
waveform will be seen on your oscilloscope.
Output
Configuration
Commands
The output configuration commands control the output function,
shape, frequency, amplitude, filter and state. Optional modes
are omitted from these commands.
Selecting an
Output Function
Type
Use the following command to select the output function type:
FUNCtion:MODE {FIXed | USER | SEQuence | SWEep}
When "FIXed" is selected, the synthesizer outputs the standard
waveform currently selected by the FUNC:SHAP command.
When "USER" is selected, the synthesizer outputs the arbitrary
waveform currently selected by the TRAC:SEL command. When
"SEQuence" is selected, the synthesizer outputs the sequence
that is programmed using the SEQ:DEF command. When
“SWEep” is selected, a sweep waveform is calculated according
to the sweep settings and downloaded into the instrument. See
Using Sweep Mode
for more information.
CAUTION
Placing the 3152A in Sweep Mode overwrites any
waveforms, segments or sequences that are currently
stored in 3152A memory. Be sure to back up any critical
data stored in the 3152A before entering Sweep Mode.
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...