3152A User Manual
Racal Instruments
©
2001
Using The Instrument 3-11
Programming the
Output Amplitude
and Offset
Use the following command to program the peak-to-peak
amplitude for the generated waveform.
VOLTage
{<amplitude>|MINimum|MAXimum}
MIN selects the smallest amplitude. MAX selects the largest
amplitude. The default amplitude is 5.00 V (into 50
Ω
).
The query:
VOLTage?
queries the output amplitude for the currently selected function
and returns a value in volts.
The command:
VOLTage:OFFSet
<offset>
sets the offset for the currently active function. The default offset
is 0 V.
The query:
VOLTage:OFFSet?
queries the output offset for the currently selected function and
returns a value in volts.
Selecting the
Filter Type
Before selecting the filter type, use the following command to
activate the filter:
OUTPut: FILTer { OFF | ON}
ON enables the filter that has been selected with the
OUTP:FILT:FREQ command. The default filter state setting is
OFF.
The query:
OUTPut:FILTer?
queries the output filter state and returns "0" (OFF) or "1" (ON).
The command:
OUTPut:FILTer:FREQuency {<20MHz | 25MHz |
50MHz>}