3152A User Manual
Racal Instruments
©
2001
3152A Specifications A-5
For trigger operation when synchronized, the trigger circuit in
the master module has control of the sample clock for all
modules. Therefore, the master module must be triggered
before any slave can output it's waveform.
Backplane Multi-
Module
Synchronization
Multiple Model 3152A modules may be synchronized. A master
Model 3152A module provides the necessary signals to slave
Model 3152A modules to achieve synchronization. The signals
required are sample clock (distributed on ECLTRG0) and
another signal (distributed on TTLTRG1). Phase offset is
settable (in degrees) for each slave module with respect to the
master module output.
Phase Offset Resolution (P.O.R.)
360
°
/ Number of Points
(min.
#
points
=
10)
Phase Offset Range
0
°
to 360
°
- 2880/n, where
n
=
number
of
points
Phase Accuracy
±[( 20 nS x 360
°
/PERIOD)
+ P.O.R.]
P.O.R.
=
phase
offset
resolution
PERIOD
=
1/(output
signal
frequency)
Accuracy
specified
after
PHAS:LOCK:NULL
command
Jitter
None
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...