3152A User Manual
SCPI and HS Command Reference 4-22 Racal Instruments
©
2001
*IDN?
- Query the synthesizer’s identity. The returned data is
organized into four fields, separated by commas. The
synthesizer responds with its manufacturer and model number
in the first two fields, and may also report its serial number and
options in fields three and four. If the latter information is not
available, the device must return an ASCII 0 for each. For
example, Model 3152A’s response to *IDN? is:
RACAL
INSTRUMENTS,3152A,0,
x.x
.
Where
x.x
is the current firmware revision
*OPC
- Set the "operation complete" bit (bit 0) in the Standard
Event register after the previous commands have been
executed.
*OPC?
- Returns "1" to the output buffer after all the previous
commands have been executed. *OPC? is used for
synchronization between a controller and the instrument using
the MAV bit in the Status Byte or a read of the Output Queue.
The *OPC? query does not affect the OPC Event bit in the
Standard Event Status Register (ESR). Reading the response
to the
*OPC?
query has the advantage of removing the
complication of dealing with service requests and multiple polls
to the instrument. However, both the system bus and the
controller handshake are in a temporary hold-off state while
the controller is waiting to read the *OPC? query response.
*OPT?
- Returns “1” if the Model has 512k of memory
installed.
*RST
- Resets the synthesizer to its default state. Default
values are listed in
Table 4-1.
*SRE
<enable_value> - Enables bits in the Status Byte enable
register.
*SRE?
- Query the Status Byte enable register. The
synthesizer returns a decimal value in the range of 0 to 63 or
128 to 191 since bit 6 (RSQ) cannot be set. The binary-
weighted sum of the number represents the value of the bits of
the Service Request enable register.
*STB?
- Query the Status Byte summary register. The *STB?
command is similar to a serial poll but is processed like any
other instrument command. The *STB? command returns the
same result as a serial poll, but the "request service" bit (bit 6)
is not cleared if a serial poll has occurred.
*TRG
- Triggers the synthesizer from the remote interface. This
command effects the synthesizer if it is first placed in the
Trigger or Burst mode of operation and the trigger source is set
to "BUS".
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...