Using The Instrument 3-25
Selecting the
Trigger Level
The trigger level command sets the threshold level at the trigger
input connector. Trigger levels are adjustable from –10 V to +10
V with up to 4 digits. Trigger level resolution is always 10 mV.
Use the following command to set the trigger level for the trigger
signal:
TRIGger:LEVel
<value>
This sets the trigger level for the signal which is applied to the
TRIG IN. The default value is 1.6 V so there is no need to modify
this value if you apply a TTL level signal to the trigger input.
TRIGger:LEVel?
queries the trigger level and returns a value in Volts.
Using the Soft
Trigger
There are a number of commands that are available to trigger
the Model 3152A. The soft trigger command is one of them. To
use the soft trigger command, place the instrument in the
TRIG:SOUR EXT mode. Soft trigger is ignored in the internal or
TTLTrg<n> modes. Use the following SCPI commands to trigger
the instrument:
TRIGger
The IEEE-STD-488.2 common command *TRG will have the
same effect. Use either software command to trigger the Model
3152A in Trigger, Burst and Triggered Sequence Advance
modes.
Generating
Arbitrary
Waveforms
The Model 3152A cannot generate arbitrary waveforms without
first loading them into memory. A description of the arbitrary
waveform function and an explanation on how to load waveforms
into memory is given in the following paragraphs.
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...