3152A User Manual
SCPI and HS Command Reference 4-30 Racal Instruments
©
2001
-200,"Execution error". This is the generic syntax error for the
instrument when it cannot detect more specific errors.
Execution error as defined in IEEE-488.2 has occurred.
-221,"Setting conflict". Two conflicting parameters were
received which cannot be
executed without generating an error.
An offset value which is outside the amplitude level window may
have been sent.
-222,"Data out of range". Parameter data which followed a
specific header could not be used because its value is outside
the valid range defined by the synthesizer.
224,"Illegal parameter value". A discrete parameter was
received which was not a valid choice for the command. An
invalid parameter choice may have been used.
-300,"Device-specific-error". This is the generic device-
dependent error for the instrument when it cannot detect more
specific errors. A device- specific error as defined in IEEE-
488.2 has occurred.
-311,"Memory error". Indicates that an error was detected in the
instrument’s memory.
-350,"Too many errors". The error queue is full because more
than 30 errors have occurred. No additional errors are stored
until the errors from the queue are removed. The error queue is
cleared when power has been shut off, or after a *CLS
command has been executed.
-410,"Query INTERRUPTED". A command was received which
sends data to the output buffer, but the output buffer contained
data from a previous command (the previous data is not
overwritten). The output buffer is cleared when power is shut off
or after a device clear has been executed.
Device-Specific
Commands
The Model 3152A conforms to the 1993.0 version of the SCPI
standard. Some of the commands used are not included in the
1993.0 version. However, these commands are designed with
the SCPI standard in mind and they follow all of the command
syntax rules defined by the standard.
Table 4-1
lists all device-
specific commands that were designed specifically for the
Model 3152A as not confirmed SCPI 1993.0 commands.
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...