3152A User Manual
Maintenance And Performance Checks 5-10 Racal Instruments
©
2001
Using WaveCAD, change the axis (Setup | Axis) to have 4000
points (0 to 3999) for the first 4 steps below and 1000 points (0
to 999) for step 5 below. Download a 1k or 4k sine waveform
(Wave | Sine | OK) into segment 1.
FREQ:RAST # of Points Frequency Distortion Reading
40e3
4000
10Hz
< 0.1%
40e4
4000
100Hz
< 0.1%
40e5
4000
1kHz
< 0.1%
40e6
4000
10kHz
< 0.1%
100e6 1000
100kHz
< 0.1%
Tune the spectrum analyzer to the carrier frequency and adjust
the gain so the fundamental corresponds to 0 dB.
Connect the Model 3152A output to spectrum analyzer input
through a 20dB feedthrough attenuator.
Set the Model 3152A FREQ setting and verify the harmonic
distortions levels as follows:
FREQ SETTING
HARMONICS LEVEL
5.000
MHz
<48
dB
50.00
MHz
<28
dB
Sine Flatness
Level Flatness
:
±
1% to 1.000 MHz;
±
5% to 10.00 MHz;
±
15%
to 50.00 MHz
Equipment
:
Oscilloscope
1.
Configure the Model 3152A as follows:
Function:
Sine
wave
Frequency:
1.000
KHz
Amplitude:
6
V
Remote
Commands:
:RES
:OUTP ON
:FREQ
10e3
:VOLT 6