3152A User Manual
Racal Instruments
©
2001
SCPI and HS Command Reference 4-19
Trigger Command
Summary
The Trigger commands control the trigger modes of the Model
3152A. The Model 3152A can be placed in Triggered, Gated or
Burst mode. Trigger source is selectable from an external
source, internal trigger generator, backplane TTLTrg 0-7, and
software trigger. Optional modes are omitted from these
commands. Factory defaults after *RST are shown in bold
typeface. Parameter low and high limits are given where
applicable.
Command and Parameters
Low Limit
High Limit
Default
INITiate:CONTinuous
{OFF
|
ON}
ON
TRIGger:BURSt
{OFF
|
ON}
OFF
TRIGger:COUNt <value>
1
1e6
1
TRIGger:DELay <value>
0/10
2e6
0
TRIGger:DELay:STATe
{OFF
|
ON}
OFF
TRIGger:GATE
{OFF
|
ON}
OFF
TRIGger:LEVel <value>
-10
10
1.6
TRIGger:SLOPe
{POSitive
|
NEGative}
POS
TRIGger:SOURce:ADVance {EXTernal | INTernal | TTLTrg<n>}
EXT
TRIGger:TIMer: <value>
15e-6
1000
100e-6
TRIGger:IMMediate
*TRG
Backplane Inter-
Module Phase
Synchronization
Command
Summary
Phase Synchronization commands control the phase offset
between two or more modules. There is no limit on how many
modules can be synchronized, as long as one module is
programmed to be master and the rest of the modules are
slaves. The location of the slave modules in relation to the
master module does not affect the accuracy of the phase offset.
The commands are presented exactly as they should be typed in
your program. Optional nodes were omitted from these
commands. Factory defaults after *RST or front panel reset are
shown in
bold typeface
. Parameter low and high limits are given
where applicable.
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...