3152A User Manual
Racal Instruments
©
2001
Maintenance and Performance Checks 5-7
Remote Commands: :RES
:OUTP ON
:FUNC:SHAP
SIN
:FREQ
1000
:VOLT
15
2.
Set the DMM to ACV measurements (RMS).
3.
Connect the Model 3152A output to DMM input.
Terminate the output with a 50
Ω
feedthrough
termination. Set the amplitude and verify DMM reading
as follows:
3152A
SETTING
DMM
READING
15.00 V
5.373 V - 5.230 V
1.500 V
537.3 mV - 523.0 mV
150.0 mV
53.73 mV - 52.30 mV
DC Offset
Characteristics
Accuracy Specifications
:
±
(1% of r1% of amp200
μ
V),
∀
80 mV window
±
(1% of r1% of amp2 mV),
∀
800 mV window
±
(1% of r1% of amp20 mV)
∀
8 V window
Equipment
:
DMM, 50
Ω
feedthrough termination
1.
Configure the Model 3152A as follows:
Function:
Sine
Frequency:
1.000 MHz
Amplitude:
1.61 V
Remote Commands: :RES
:OUTP ON
:FUNC SIN
:FREQ 1e6
:VOLT 1.61
:VOLT:OFFS 5