3152A User Manual
Maintenance And Performance Checks 5-26 Racal Instruments
©
2001
Table 5-4, Sequence Generator Checkout Procedure
Step
Test Point
Expected Result
1
U45 pin 1, U42 pin 11
Negative going pulses; 2 ns width, 100
μ
s period
2
U46 pin 1, and 3
Same as above
3
U42 pin 12
Negative going pulses; 100
μ
s width, 200
μ
s period
4
U42 pin 8, U43 pin 14
Negative going pulses; 2
μ
s width, 100
μ
s period
5
U44 pin 14, U47 pin 14
Same as above
6
U47 pin 5
Negative going pulses; 2
μ
s width, alternating 100
μ
s
and 200
μ
s periods
7
U47 pin 2, and 3
Positive going pulses; 100
μ
s width, 200
μ
s period
The next checkout is done on the Model 3152A configured to
operate in burst mode with 5 bursts. Use the following
command to configure the Model 3152A:
Remote Commands:
:FUNC:MODE USER
:TRAC:SEL 1
:INIT:CONT
OFF
:TRIG:BURS: ON
:TRIG:COUN 5
:OUTP ON
Connect an external trigger source to the front panel TRIG IN
connector. Set external trigger frequency to 1 KHz. Proceed
with the checkout procedure given in
Table 5-5
.